Trio™ Operating Instructions 0070-10-0666-01 2 - 61
Operations Parameter Menus
IBP Alarm Limits
Alarms occurring during the process of IBP measurement include two (2) types: physiological
alarms and technical alarms. Physiological alarms occur when the patient’s pressure values
are equal to or exceed set alarm limits. Technical alarms are any IBP-related alarms, which
are not physiological, such as functional failures.
IBP
LABEL
SYS
HIGH
(mmHg)
SYS
LOW
(mmHg)
MEAN
HIGH
(mmHg)
MEAN
LOW
(mmHg)
DIA
HIGH
(mmHg)
DIA
LOW
(mmHg)
ART
Adult
5 – 300
[180]
0 – 150
[80]
5 – 150
[100]
- 5 – 100
[40]
5 – 140
[100]
0 – 120
[50]
Ped
5 – 240
[150]
0 – 130
[70]
5 – 100
[80]
- 5 – 50
[30]
5 – 100
[80]
0 – 95
[40]
PA
Adult
5 – 300
[180]
0 – 150
[80]
5 – 150
[100]
- 5 – 100
[40]
5 – 140
[100]
0 – 120
[50]
Ped
5 – 240
[150]
0 – 130
[70]
5 – 100
[80]
- 5 – 50
[30]
5 – 100
[80]
0 – 98
[40]
CVP
Adult
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
RAP
Adult
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
LAP
Adult
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped
n
/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
ICP
Adult
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
* Factory default values shown in brackets.
*