Mindray TRIO Instruções de Operação Página 166

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 165
Trio™ Operating Instructions 0070-10-0666-01 6 - 27
Appendix Physical Specifications
Isolation: Risk (Leakage) Currents:
Meets the requirements of IEC 60601-1/
EN 60601-1
Enclosure Risk Current:
Normal operating conditions <= 100 μA
Single fault condition <= 300 μA for 120 VAC
Single fault condition <= 500 μA for 230 VAC
Patient Source Current:
Normal operating conditions <= 10 μA
Single fault condition <= 50 μA
Patient Sink Current:
Normal operating conditions not applicable
Single fault condition <= 50 μA
Dielectric Withstand Per IEC 60601-1/EN 60601-1