C-3
NIBP COMM ERR
Yes Yes H Communication
between NIBP
module and the host
fails.
Restart the monitor. If
the problem still
exists, contact our
service personnel for
repair.
LOOSE CUFF No Yes L The NIBP cuff is not
properly connected.
AIR LEAK No Yes L
CUFF TYPE ERR No Yes M
PNEUMATIC
LEAK
No Yes L
Leak in the airway.
AIR PRESSURE
ERROR
No Yes L
NIBP WEAK
SIGNAL
No Yes L
SIGNAL
SATURATED
No Yes L
Check the patient’s
condition and verify
patient type. Replace
with a appropriate
cuff and connect it
correctly. If the
problem still exists,
contact our service
personnel for repair.
NIBP
OVERRANGE
No Yes M
Failures occur in the
pulse measurement.
The monitor cannot
perform
measurement,
analysis, or
calculation. 。
EXCESSIVE
MOTION
No Yes L Excessive motion of
the patient’s arms
OVER
PRESSURE
No Yes M The airway might be
blocked.
NIBP SYSTEM
FAILURE
No Yes M
NIBP TIME OUT No Yes M
Failures occur in the
pulse measurement.
The monitor cannot
perform
measurement,
analysis, or
calculation.
Check the patient’s
condition and verify
patient type. Replace
with a appropriate
cuff and connect it
correctly. If the
problem still exists,
contact our service
personnel for repair.
NIBP RESET
ERROR
No Yes M Illegal reset during
NIBP measurement.
Check the airway and
take measurements
again. If the problem
still exists, contact
our service personnel
for repair.