4-3
In the following circumstance, the setting of ALARM VOL may be lower than the
setting of MIN ALARM VOL. In this case, ALARM VOL is automatically adjusted
according to MIN ALARM VOL.
Select SYSTEM SETUPMAINTAINUSER MAINTAIN; enter required
password; and then change MIN ALARM VOL in the pop-up menu.
Select SYSTEM SETUPDEFAULT; and then select LOAD FACTORY
CONFIG or LOAD USER CONFIG in the pop-up menu;
Restart the monitor after turning it off for more than 120s;
Change patient ID;
Change patient type.
KEY VOL 0 to 10
PULSE VOL 0 to 10
LCD BRIGHT 1 to 10
LCD CONTRAST 1 to 10
NIBP UNIT mmHg, kPa
When the key volume or pulse volume is set to 0, it indicates that the key tone or pulse tone
is disabled.
4.2.2 Network Setup
In the SYSTEM SETUP menu, select NET SETUP. In the NET SETUP menu, you can set
the following items.
NET TYPE: CMS, CMS+
LOCAL NET NO.:It indicates the bed number of a monitor in the monitoring network.
If the NET TYPE is CMS, the LOCAL NET NO can be set between 1 and 64; if the
NET TYPE is CMS+, it can not be set.
IP ADDRESS SETUP: When the monitor is connected with the central monitoring
system, and the NET TYPE is CMS+, you can set the IP address for your monitor .
The network type and local net No. are related to the central monitoring system (CMS) to
which the monitor is connected. Contact the CMS technical personnel for any doubt.
WARNING
This monitor is intended for use only by clinical professionals or under their
guidance. It must only be used by persons who have received adequate training
in its use. Anyone unauthorized or untrained must not perform any operation
on it.