Vista de página 134
10-31
In the 12-Lead Monitoring Display:
Labels except the ECG-related ones (ECG, ST, GAIN, RHYTHM) are all disabled.
The REC button on the control panel is disabled.
You can press the MAIN and MENU buttons on the control panel to exit the 12-Lead
Monitoring Display and return to the normal monitoring display.
10.7.2 12-lead ST Analysis Menu
Select the ST label in the parameter area, or select ST ANALYSIS >> in the ECG SETUP menu.
The following menu appears.
Figure 10-22 ST ANALYSIS menu
In the ST ANALYSIS menu, you can perform the following settings:
ST ANAL ST analysis
ON: Enables the ST analysis;
OFF: Disables the ST analysis.
NOTE
z When ST ANAL is set to ON, the monitor selects DIAGNOSTIC mode
automatically. You can set the monitor to MONITOR or SURGERY mode as
required. However, ST analysis will turn off automatically.